IAQ Monitor FAQ's
How long can the aq-5000/5001 or AQ5000/5001Pro run before memory is
full?
| # OF CHANNELS ON |
INTERVAL |
RUN TIME |
| 1 |
1 Second |
15 hr. 48 min. |
| 2 |
1 Second |
7 hr. 54 min. |
| 3 |
1 Second |
5 hr. 15 min. |
| 4 |
1 Second |
3 hr. 57 min. |
| 5 |
1 Second |
3 hr. 9 min. |
|
|
|
|
| 1 |
1 Minute |
39 days 13 hr. |
| 2 |
1 Minute |
19 days 18 hr. |
| 3 |
1 Minute |
13 days 4 hr. |
| 4 |
1 Minute |
9 days 19 hr. |
| 5 |
1 Minute |
7 days 21 hr. |
|
|
|
|
| 1 |
5 Minute |
197 days 21 hr. |
| 2 |
5 Minute |
98 days 22 hr. |
| 3 |
5 Minute |
65 days 19 hr. |
| 4 |
5 Minute |
49 days 9 hr. |
| 5 |
5 Minute |
39 days 13 hr. |
| |
|
|
| 1 |
10 Minute |
395 days 19 hr. |
| 2 |
10 Minute |
197 days 21 hr. |
| 3 |
10 Minute |
131 days 14 hr. |
| 4 |
10 Minute |
98 days 18 hr. |
| 5 |
10 Minute |
79 days 3 hr. |
| |
|
|
| 1 |
15 Minute |
593 days 18 hr. |
| 2 |
15 Minute |
296 days 21 hr. |
| 3 |
15 Minute |
197 days 9 hr. |
| 4 |
15 Minute |
148 days 3 hr. |
| 5 |
15 Minute |
118 days 18 hr. |
Do the temperature and humidity sensors on my
indoor air quality monitor need to be calibrated?
There is no field calibration needed for the
temperature and humidity sensor, however these sensors will undergo
verification testing when your aq-5000/5001 is sent in for factory recertification.
How long can the AQ-5000/5001 record?
The aq-5000/5001 has enough memory to store
57,344 data points. For example, with 3 channels (CO2, Temp and
Humidity) recording data points every minute the unit can record for 13
days and 6 hours.
What are the recommended calibration gases for
the CO2 sensor ?
To zero the CO2 sensor you should use Nitrogen
(N2). For the span of the CO2 sensor you need to consider what range you
want to monitor. The aq-5000/5001 CO2 channel can be programmed for
either 5000 or 20,000 ppm full scale. For the 5000 ppm full scale range
we recommend using 1000 or 5000 ppm CO2. For the 20,000 ppm full scale
range we recommend using 10,000 ppm CO2.
How do I set the date and time on my
aq-5000/5001 ?
The date and time of the aq-5000/5001 are set
automatically when the unit is programmed using the software package. Using the
software, simply create a new or view an existing setup file, program
the recorder and the date and time from the computers real time clock
are sent to the aq-5000/5001. The aq-5000/5001 date and time will now
match the computer from which it was programmed.
I just finished calibrating the CO2 sensor on my
aq-5000/5001 and the reading is very unstable, what could be wrong?
In most instances unstable readings are a direct
result of the user calibration. In some instances users have the
incorrect gas attached to the unit when they press the enter button. In
other instances they have set the CO2 span level to a number that does
not match the CO2 span gas being used. In any instance the unstable
reading should be corrected by simply performing another calibration
being sure to do the CO2 zero first and then the CO2 span. Because of
interaction between the zero and span it may be necessary to zero and
span the unit and then zero and span the unit a second time. This will
usually return the reading to a stable condition.
How does the aq-5000 Pro Calculate Dewpoint?
Dewpoint = eighth_root( RH/100 ) * ( 9*Temp/10 + 112 ) - 112 + Tc/10
where:
RH is the relative humidity in percent
Temp is the temperature in degrees Celsius
What is STEL?
STEL is a 15 minute rolling average (that rolls each minute). This
means that the average reading is calculated after the first 15 minutes of
logging and then again after each subsequent minute of logging. At the
completion of each subsequent minute, the new 15-minute average is
compared with the old and the higher of the two is stored and displayed.
What is TWA?
TWA is an average based on an 8 hour interval. The average is taken by
summing all of the the samples taken and dividing by the number of samples
in 8 hours (28800 for the aq-5000 Pro). The important thing here is that
the average is calculated the same way regardless of whether you've run
the instrument for 1 hour, 8 hours, 24 hours or any other amount of time
(i.e. the number of samples used to divide by is the same).